Specifications∶
  ·Parameters: WBC, LYM#, MID#, GRAN#, LYM%,
MID%, GRAN%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-CV, RDW-SD, PLT, MPV, PDW,
PCT,P-LCR,P-LCC
  3-part differentiation of WBC, 21
parameters and 3 histograms
  ·Principles of operation:WBC/RBC/PLT:
Electrical Impedance
                HGB: Photoelectric
colorimetry
  ·Aspiration Volume:Whole blood
10uL,Pre-diluted 20uL
  ·Aperture Diameter:WBC 100um,RBC/PLT 68um
  ·Throughput:60 samples per hour
  ·Data storage:More than 100,000 sample
results with histograms
  ·Alarms: Error messages
  ·Dilution Ration:               Whole blood       Capillary blood
          WBC/HGB        1:232            1:400
         RBC/PLT        1:40000           1:45000
  ·Display: 10.4 inches color LCD
  ·Control and calibration: L-J,X,X-R,X-B
four control modes,manual and auto-Calibration System check
  ·Dimension: 330mm x 440mm x 500mm
  ·Power requirement: AC 100V --- 240V,
50/60Hz, 180W
  ·Input/Output: Support RS232, standard
network port and USB
          Shortcut keys with standard mouse
and keyboard
          Internal 57mm thermal printer,
          Parallel Port and USB for external
printer
Characteristics:
  ·Large color LCD display & Linux
operation system
  ·Support USB data backup & system
upgrade
  ·Excellent data management with mouse and
keyboard
  ·High efficient self checking system, low
maintenance
  ·Support LIS and HIS with HL7 protocol
  ·Collaborator of BIO-RADcontrol,target
value available
|